Add to Cart
DSC-800B Differential Scanning Calorimeter/Oxidation induction time
Technical feature:
1.Industrial-grade widescreen touch structure with comprehensive display: set temperature, sample temperature, oxygen flow rate, nitrogen flow, differential thermal signals, various switch states, and flow zero indication.
2.USB communication interface with high universality, reliable non-interruptive connection, and support for recovery function.
3.Compact furnace design with infinitely adjustable cooling rate.
4.Enhanced installation process using mechanical fixing to completely prevent internal colloid from contaminating differential thermal signals.
5.Dual temperature probes ensure repeatable sample temperature measurement at specified heights.
6.One temperature sensor on the furnace wall enables PID control of furnace temperature, but thermal inertia causes sample temperature deviation (varying by season). Single-probe measurement leads to significant errors in both thermal and temperature signals. The instrument installs a bottom-mounted probe to measure true sample temperature, using proprietary control technology to adjust furnace wall temperature to the set value.
7.Digital gas mass flowmeter for automatic two-way atmosphere switching with fast response and short stabilization time.
8.Calibrated with standard samples for easy customer adjustment of temperature coefficients.
9.Software adapts to all screen resolutions, automatically optimizing curve display based on screen size. Compatible with laptops/desktops, supporting Win2000, XP, VISTA, Windows 7, etc.
10.User-programmable procedures enable fully automated measurements. Dozens of commands allow flexible combination and saving, simplifying complex operations to one-key functions.
Temperature range | room temperature ~800℃ |
Temperature resolution | 0.01℃ |
Heating rate | 0.1~80℃/min ; |
Temperature control mode | rise in temperature, constant temperature (Programs control automatically) |
DSC range | 0~±200mW |
DSC resolution | 0.01mW |
DSC sensitivity | 0.1mW |
Working power supply | AC 220V 50Hz or custom-made |
Atmosphere control of gas | nitrogen ;oxygen (instrument changes automatically) |
Gas flow | 0-200mL/min |
Gas pressure | 0.2MPa |
Accuracy of gas flow | 0.2mL/min |
Matching crucible | Alumina crucible Φ6.7*3mm |
Parameter standard | equipped with standard material(indium ,tin, zinc) the user can correct the temperature and enthalpy |
Data interface | standard USB interface |
Display mode | 24bit color 7 inch LED touch screen display |
Output mode | computer and printer |
Model | quantity | unit | note |
DSC | 1 | SET | Fit Model: DSC-500A DSC-500B DSC-800B DSC-500BL DSC-1150B |
Aluminum crucible | 300 | PCS | |
Standard test samples (indium, tin, and zinc) | 1 | Part | |
power wire and USB | 1 | PCS | |
Software CD | 1 | PCS | |
tweezers | 1 | PCS | |
Medicine spoon | 1 | PCS | |
Custom valve connector | 2 | PCS | |
Oxygen and nitrogen gas duct | 5 | Meter | |
quick coupling | 2 | PCS | |
Soft dog | 1 | PCS | |
Fusing glass fuse | 4 | PCS |